Real Time PCR System, PCR-Q96 Series
- Sample capacity: 96 holes
- Fluorescent channels: 4/6
- Max. Heating/Cooling rate: 6.0° C/s
- No ROX correction required.
- New optical scanning detection system
- Other names: Rapid PCR machine, RT-PCR machine, QPCR machine
Description
As a necessary choice for quantitative analysis of molecular biology, Real-time PCR system has been widely used in various fields such as scientific research, clinical detection and diagnosis, quality and safety testing, and forensic applications.
- Up to 6 fluorescence detection channels allowing multiplex PCR.
- Effectively reduce multi-color crosstalk and edge effect, no ROX correction required.
- New optical scanning detection system
- Innovative scanning method and time-resolved signal separation technology
- Unique edge temperature compensation technology
Specifications
Model | PCR-Q96-5 |
PCR-Q96-6 |
Sample capacity | 96 | |
Reaction volume | 10-50 μl | |
Thermal cycle technology | Peltier | |
Temperature control system | ||
Max. Heating/Cooling rate | 6.0° C/s | |
Heating temperature range | 4 – 100 °C | |
Temperature accuracy | ± 0.2°C | |
Temperature uniformity | ±0.2℃ @60℃ , ±0.3℃ @95℃ | |
Temperature gradient setting range | 30–100°C | |
Temperature gradient difference setting range | 1 – 36°C | |
Detection system | ||
Excitation light source | 4 monochrome high efficiency LEDs | 6 monochrome high efficiency LEDs |
Detection device | PMT | |
Detection mode | Time-resolved signal separating technology | |
Excitation/detection wavelength range | 455-650nm/510-715nm | |
Fluorescent channels | 4 channels | 6 channels |
Supported dye | FAM/SYBR Green, VIC/JOE/HEX/TET, ABY/NED/TAMRA/Cy3, JUN, ROX/Texas Red, Mustang Purple, Cy5/LIZ | |
Sensitivity | Single copy gene | |
Resolution | 1.33 folds copy number difference can be distinguished in single-plex qPCR | |
Dynamic range | 10 orders of magnitude copies | |
Package Dimension (W*D*H)(mm) | 960*660*750 | |
G.W.(kg) | 47 |